The components of an anesthesia machine work together to deliver controlled amounts of oxygen and anesthetic gas.
• Components work together to deliver controlled amounts of oxygen and anesthetic gas
• Everything after the common gas outlet
o Anesthetic tubing
o Rebreathing bag
o CO2 absorber
o Scavenge system
• To deliver oxygen and anesthetic gas
• To remove CO2 from exhalation
• To facilitate controlled ventilation
• Circle system or Rebreathing system
o Commonly used in patients greater than 7 kg
• Utilizes inspiratory and expiratory valves as well as CO2 canister
o Closed or partially closed system
• Also called low-flow anesthesia
• Refers to the oxygen flow rate and position of the pop-off valve
• Non-rebreathing system
o Most often used for patients less than 7 kg
• Anesthesia machines are designed to deliver volatile anesthetics in the presence of oxygen: Oxygen is the carrier gas for anesthetic
o 100% oxygen is commonly used to deliver anesthetics
o This is necessary for the delivery of adequate amounts of O2 to tissues
• Anesthetics reduce tidal volume
• Anesthetics cause some degree of respiratory depression
• It is important to avoid hypoxemia
• Low arterial oxygen concentration
• Obtained as compressed gas
• Pressurization is necessary to fit a large amount into a relatively small container
• Sizes A-H available, most common are sizes E and H
• Full cylinders contain 2200 psi of gas with varying volumes based on size
o E-cylinders
• Small and easy to carry
• Holds ~ 770 L of oxygen
• Commonly found attached to an anesthetic machine
o H-cylinders
• Huge and heavy
• Holds ~ 7000 L of oxygen
• Commonly found in a closet or chained to a wall
• All medical gas cylinders are color coded for safety
o Oxygen is green
o Nitrous oxide is blue
o Air is yellow
o Nitrogen is black
• These colors are true in the US only
• Provides a safe, constant operating pressure within the machine regardless of the pressure in the tank
• Typically expressed in L/min
• Oxygen enters the bottom of the flowmeter and exits the top
• Allows the anesthetist to adjust the O2 flow rate
• The flow of oxygen can be increased to speed the change of inhalant concentration in the machine
• A time constant represents the volume of the machine in relation to the flow of gases
• It takes approximately 3 time constants to see a 95% change in the concentration within the system
• Machine volume (components of breathing system)
o Sodasord canister ~ 1 L
o Rebreathing bag- depends on size
o Breathing hoses- depends on length and diameter
o Arbitrary example volume 5 L
• Math
o 3 TC X 5 L / 1L/min = 15 min With a flowrate of 1 L/min it takes 15 min to approach a steady state
o 3 TC X 5L / 2L/min = 7.5 min With a flowrate of 2 L/min it takes 7.5 min to approach a steady state
• Factors affecting changes in anesthetic concentration
o Hypoventilation and apnea
• By assisting ventilation you facilitate gas exchange
• This includes oxygen as well as anesthetic
• CAUTiON
o Increases in flowrates should always be monitored closely
o An increase in system flow also means an increase in flow to patient
• Can be very drying to respiratory tract
• Can cool patients very quickly
o Never close the pop-off valve when you have higher flows
• Pressure builds much quicker so you have less time to react!
o High flows waste anesthetic gas and oxygen and money!
• Convert liquid anesthetic to gas based on vapor pressure
• Adds controlled amounts of anesthetic gas to O2
• Can only release anesthetic in the presence of oxygen (or N2O)
• Each vaporizer is calibrated for a specific agent based on that agents vapor pressure
• Never fill a vaporizer with a different anesthetic
• Allows fresh oxygen to by-pass vaporizer and enter breathing circuit
• Delivered directly at 30-50 L/min
• Handy but dangerous
• Over pressurization is easy
• Use with extreme care or not at all when the circuit is connected to a patient
• NEVER use with a non-rebreathing system
• NEVER use when pop-off is closed
• DO NOT use in small patients
• The use of this valve decreases anesthetic concentration in the system
• Gas mixture (O2 and anesthetic) exits vaporizer and machine at the fresh gas outlet
• The mixture follows a flexible tube and enters the breathing system at the fresh gas inlet
o This is necessary to hook up a bain non-rebreathing circuit
o These hoses are internal on some machines
• Fresh gas enters the one-way inspiratory valve on its way to the patient and then is exhaled through the one-way expiratory valve
• The Y-piece helps facilitate the one-way gas flow
• Only present in the circle system
• Allows for re-breathing of gases
• Should hold a minimum of 60 ml/kg of the patients body weight
o 5-6 X tidal volume (10-15 mL/kg)
o Always round up
• Allows patient to take a large breath
• Allows anesthetist to observe breathing
• Allows anesthetist to breathe for the patient
• Size of the bag does matter
o If bag is too big
• Impairs monitoring of breathing rate
• Adds volume to machine
• Slows changes in inspiratory anesthetic concentrations when settings are altered
o If bag is too small
• Animal is unable to take an adequate breath
• Bag collapses on inspiration
• Transports anesthetic gas from machine to patient
• At least two different diameter hoses available for small animals
o 22 mm for larger dogs
o 15 mm for small dogs and cats
• Different lengths also available
• Many are one-time use only in humans
o Cuffs tend to wear out over time
• All ET tubes should be checked for leaks prior to use
o Make sure cuff inflates
o Can place inflated tube in a bowl of water and look for leaks/bubbles
o Discards any leaky tubes
• Prevents build-up of pressure or volume within the circuit
• Vents pressure at 2 cm of water
• Pressure build-up may impede venous return or cause barotrauma
o The distention and possible rupture of alveoli from excessive pressure which may lead to pneumothorax and/or subQ emphysema
• Always keep a hand on the pop-off valve (when you close it) to make sure that it gets re-opened
• Contains absorbent granules that remove CO2 from gas
• Any gas that gets returned to the patient passes through here on its way back
• It lowers fresh gas flows
• Reduces waste of anesthetic and oxygen
• Lowers the overall cost of anesthesia
• Allows for re-breathing of gases
• Calcium hydroxide or barium hydroxide
• Exothermic reaction
o Produces both heat and water
• Must be changed when exhausted
o Won't absorb CO2 if spent
o May lead to patient hypercarbia if spent
o Has a pH color indicator to tell when to change it
• White to lavender or blue
• Short-lived chemical reaction; will change back to white even if spent
o Fresh granules can be easily chipped and crumbled
o Saturated granules are hard and brittle
• Measures the pressure of gases within the anesthetic circuit and patient
o Useful when breathing for a patient or setting up a ventilator
• Pressure should not exceed 15-20 cm H2O or 10-14 mmHg
• Collects waste gas and disposes of it
o Passive disposal
• Non-circulating ventilation systems
• Activated charcoal
o Active disposal
• No exhaled gas is returned to patient
o Adequate oxygen flows required to prevent rebreathing of gases
o 250-500 ml/kg/min
• Evacuated by scavenge based on O2 flows
• No CO2 absorber
• When to use depends on patient size
• Many types, two common are both modified Mapleson D
o Ayers T
o Bain
• Circle system
o Low flows
o Rebreating of gases
o More economical
o Recycled air is warmer and humidified than fresh gas
o May be cumbersome
o More resistance to breathing for small patients
• Non-rebreathing system
o Little resistance to breathing
o Does not require a CO2 absorber
o Allows inspired concentrations of anesthetic to be changed rapidly
o Promotes hypothermia and drying of respiratory tract
o Can be wasteful if used in larger patients (high flows)
• The area where bi-directional flow takes place
o The endotracheal tube
o Anything between the Y-piece and the ET tube
• CO2 monitor adapter, swivel adapter, elbow
• Excessive pieces should be kept to a minimum in small patients
o Hose length does not contribute to dead space but can increase the resistance to breathing
• Areas prone to leaks
o Around the sodasorb canister, around valve caps, re-breathing bag, hoses, any joint that gas passes through
• If you suspect a leak...
o Go through and tighten all joints and connections
o Leak check again
• If still leaking keep O2 on and spray a dilute soapy liquid around all joints in the breathing circuit
o The gas sneaking out of the system will cause bubble to form so keep your eyes open for them!
Dorsch JA, & Dorsch SE. Understanding Anesthesia Equipment. Baltimore: Williams & Wilkins, 1999
Hartsfield, SM. Anesthetic Machines and Breathing Systems. In: Lumb & Jones Veterinary Anesthesia and Analgesia. Iowa: Blackwell, 2007. pg. 453-494
Podcast CE: A Surgeon’s Perspective on Current Trends for the Management of Osteoarthritis, Part 1
May 17th 2024David L. Dycus, DVM, MS, CCRP, DACVS joins Adam Christman, DVM, MBA, to discuss a proactive approach to the diagnosis of osteoarthritis and the best tools for general practice.
Listen